ISO 389-7 Hearing Threshold Fill

Formula
ISO 389-7:2019 Table 1 reference threshold values (free-field or diffuse-field)
Output
dB SPL (re 20 µPa) — fills a row with the threshold of hearing
Bands
1/3-octave centres 20 Hz – 16 kHz (fills whichever of these the sheet has)
Standard / Source
ISO 389-7:2019
Access
Curve fill

What it does

Fills the selected row with the reference threshold of hearing from ISO 389-7:2019 — the lowest sound pressure level an otologically normal young listener can just detect, band by band. Useful as an overlay to show whether a quiet or low-frequency component is likely to be audible at all. Pick Curves › Hearing › Hearing Threshold; the row is labelled “ISO 389-7 free-field threshold” and a yellow dropdown cell switches between the two listening conditions:

  • Free-field (the default) — frontal incidence, pure tones.
  • Diffuse-field — one-third-octave noise bands.

ℹ️ Available on every sheet kind. Any band of your sheet that the standard does not tabulate (e.g. below 20 Hz on the low-frequency sheet) is left blank. The values are tabulated, not calculated.

Worked example — octave sheet

Band (Hz)31.5631252505001000200040008000
Free-field (dB)59.537.522.111.44.42.4−1.3−5.412.6
Diffuse-field (dB)59.537.522.111.43.80.8−1.5−3.86.8

(Below 315 Hz the two conditions are identical.)

Source

ISO 389-7:2019 — Acoustics: Reference zero for the calibration of audiometric equipment, Part 7: Reference threshold of hearing under free-field and diffuse-field listening conditions, Table 1. The 20 Hz value is 78.1 dB (the ISO 389-7 Amd 1:2016 correction of the 78.5 dB printed in ISO 226:2003).